Capacity
x Readability
(grams x µg) |
2.1
g x 0.1 µg |
| Repeatability,
micrograms
|
±
0.25 µg |
| Linearity,
micrograms |
±
0.9 µg |
| Linearity,
0 -500 micrograms |
±
0.4 µg |
| Response
Time (avg) |
10
seconds |
| Response
Time in the IQ-mode (0.01%) |
2.5
seconds |
| Tare
Range |
2.1
grams |
| Adaptation
to ambient conditions & applications requirements |
By
selection of 1 of 4 optimized filter levels |
| Display
Update |
0.2 - 0.4 seconds (depending on the filter level) |
| Operating
Temperature |
5°
C to 40 ° C (41° F to 104° F),
85% RH or less |
Sensitivity
Drift
(5°-40°C/41°F-104°F) |
±
1 ppm / °C |
| Pan
Size (mm / in) |
ø
20 mm / 0.78 in Std pan
ø 50 mm / 1.96 in Std filter |
| Chamber
Opening |
N/A |
| Electronics
Unit:WxDxH |
9.9"
x 12.6" x 4.1" |
| Deviation
of the readout when balance is tilted 1:1,000 |
±
3 µg |
| Selectable
Weight Units |
g,
kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol,
bat, MS |
|
Selectable Application Programs |
Mass
unit conversion by toggling,"EUREKA" air buoyancy correction
program, SQmin,
differential weighing and backweighing, density determination, tare memory,
net total, weighing in percent, over / under checkweighing (classification
and filling) counting |
| Auto
Zero Tracking |
Std
feature (can be turned off by menu code) |
| Built-in
Interface |
RS-232 C-S/V24-V28, RS423 /V10; 7-bit; parity; even, odd; mark, space;
Transmission rates: 150 ... 19,200 baud; 1 or 2 stop bits; software
/ hardware handshake |
|
Calibration |
Self Calibration "isoCAL" function |
|
External Calibration |
2
gram Class 0 (Optional Wgt required) |
| Power
Source |
AC Adapter 120 VAC included with scale |
| Standard
Equipment |
120V
AC Adapter, level indicator, adjustable feet, RS-232, manual |
| Optional
Accessories |
Printer,
Rechargeable external battery pack, SartoConnect data transfer software,
external calibration mass |