Capacity
x Readability
(grams x micrograms) |
31
g x 1 µg |
Repeatability
|
±
2 micrograms |
| Linearity,
micrograms |
±
10 micrograms |
| Response
Time (avg) |
14 - 18
seconds |
Response
Time
|
2.5
seconds IQ-mode (0.01%) |
| Tare
Range |
31
grams |
| Adaptation
to ambient conditions & application requirements |
By
selection of 1 of 4 optimized filter levels |
| Display
Update |
0.1 - 0.4 seconds (depending on filter level) |
| Operating
Temperature |
5°C to 40 °C (41°F to 104°F), 85% RH or less |
Sensitivity
Drift
(5°-40°C/41°F-104°F) |
±
1 ppm / °C |
| Pan
Size (mm / in) |
ø
30 mm / 1.2 in |
| Draft
Shield Ht
(mm/in) |
239
mm / 9.4 in |
| Chamber
Size (mm/in) |
223 x 410 x 232 mm (W x D x H)
8.78" x 16.14" x 9.13" (W x D x H) |
| Chamber
Opening |
approx 5 inches |
| Electronic Unit:WxDxH |
254mm x 320mm x 106mm |
| Deviation
of the readout when balance is tilted 1:1,000 |
±
5 µg |
| Selectable
Weight Units |
g,
kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol,
bat, MS |
|
Selectable Application Programs |
Mass
unit conversion by toggling,"EUREKA" air buoyancy correction
program, SQmin,
differential weighing and backweighing, density determination, tare memory,
net total, weighing in percent, over / under checkweighing (classification
and filling) counting |
| Automatic
zero tracking |
Standard
feature |
| Built-in
Interface |
RS-232 C-S/V24-V28, RS423 /V10; 7-bit; parity; even, odd; mark, space;
Transmission rates: 150 ... 19,200 baud; 1 or 2 stop bits; software
/ hardware handshake |
|
Calibration |
Self Calibration "isoCAL" function |
|
External Calibration |
30
gram Class 0 (Optional Wgt required) |
| Power
Source |
AC Adapter 120 VAC included with scale |
| Standard
Equipment |
120V
AC Adapter, level indicator, adjustable feet, RS-232, manual |
| Optional
Accessories |
Printer,
Rechargeable external battery pack, SartoConnect data transfer software,
external calibration mass |
| Warranty |
2
years manufacturer's warranty |
|
|